Title :
Achieving Commutation Control of an MRI-Powered Robot Actuator
Author :
Felfoul, Ouajdi ; Becker, Aaron ; Bergeles, Christos ; Dupont, Pierre E.
Author_Institution :
Med. Sch., Boston Children´s Hosp., Dept. of Cardiovascular Surg., Harvard Univ., Boston, MA, USA
Abstract :
Actuators that are powered, imaged, and controlled by magnetic resonance (MR) scanners could inexpensively provide wireless control of MR-guided robots. Similar to traditional electric motors, the MR scanner acts as the stator and generates propulsive torques on an actuator rotor containing one or more ferrous particles. Generating maximum motor torque while avoiding instabilities and slippage requires closed-loop control of the electromagnetic field gradients, i.e., commutation. Accurately estimating the position and velocity of the rotor is essential for high-speed control, which is a challenge due to the low refresh rate and high latency associated with MR signal acquisition. This paper proposes and demonstrates a method for closed-loop commutation based on interleaving pulse sequences for rotor imaging and rotor propulsion. This approach is shown to increase motor torque and velocity, eliminate rotor slip, and enable regulation of rotor angle. Experiments with a closed-loop MR imaging actuator produced a maximum force of 9.4 N.
Keywords :
actuators; biomedical MRI; closed loop systems; medical robotics; rotors; stators; torque control; velocity control; MR scanner; MR signal acquisition; MRI-powered robot actuator; actuator rotor; closed-loop control; commutation control; electromagnetic field gradient; ferrous particle; high-speed control; magnetic resonance scanner; motor torque; propulsive torque; rotor angle; rotor imaging; rotor propulsion; rotor slip; stator; Actuators; Commutation; Magnetic resonance imaging; Radio frequency; Rotors; Torque; MRI; Magnetic actuation; medical robots and systems;
Journal_Title :
Robotics, IEEE Transactions on
DOI :
10.1109/TRO.2015.2407795