• DocumentCode
    1776619
  • Title

    Electrical modelling and power quality analysis of three-phase X-ray apparatus

  • Author

    Moreira, Adriano C. ; da Silva, Luiz C. P. ; Morales Paredes, Helmo K.

  • Author_Institution
    Dept. of Energy & Syst, Univ. of Campinas, Campinas, Brazil
  • fYear
    2014
  • fDate
    18-20 June 2014
  • Firstpage
    1098
  • Lastpage
    1103
  • Abstract
    This article proposes the use of PSCAD/EMTDC to model and simulate a three-phase x-ray machine. The model is an alternative tool to power quality studies, load connection impact analysis and power factor correction in the electric power distribution systems. The developed model is used to identify and quantify potential impacts of the X-ray machine in the electric power distribution system. Analyzes and discussions are based on actual measurements obtained in an X-ray apparatus terminals.
  • Keywords
    X-ray apparatus; distribution networks; power factor correction; power supply quality; EMTDC; PSCAD; X-ray machine; electric power distribution systems; electrical modelling; load connection impact analysis; power factor correction; power quality analysis; three-phase X-ray apparatus terminal; Current measurement; Electron tubes; Harmonic analysis; Load modeling; Power quality; Power system harmonics; Radiography; Modelling; Non-linear loads; Power Quality; X-ray; harmonics;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Power Electronics, Electrical Drives, Automation and Motion (SPEEDAM), 2014 International Symposium on
  • Conference_Location
    Ischia
  • Type

    conf

  • DOI
    10.1109/SPEEDAM.2014.6872008
  • Filename
    6872008