  • Abstract
    An evaluation of the present approach to the design of birdcage RF coils for RMI, based on periodic lumped circuit, has been performed. It appears that, for coils bigger than λ/6, a more accurate model is required, to reduce the trial-and-error tuning of the coil.
  • Keywords
    magnetic resonance imaging; receiving antennas; transmitting antennas; RMI; high-field MRI; lumped birdcage coil model; periodic lumped circuit; trial-and-error tuning; Coils; Conductors; Inductance; Integrated circuit modeling; Magnetic fields; Magnetic resonance imaging; Resonant frequency; Birdcage Coil Model; Birdcage RF coils; MRI;