• DocumentCode
    1782163
  • Title

    Reconsideration of EMI phenomenon in active implantable medical devices in the age of MR Conditional devices

  • Author

    Toyoshima, Takeshi

  • Author_Institution
    CRDM Bus. Unit, USCI Holdings, Inc., Tokyo, Japan
  • fYear
    2014
  • fDate
    12-16 May 2014
  • Firstpage
    234
  • Lastpage
    236
  • Abstract
    Until recently, the presence of an AIMD has been considered a contraindication to MRI because of the potential adverse effects of MRI on AIMDs. However, all of AIMD manufacturers have released “MR Conditional” devices recently. These devices have been demonstrated to pose no known hazards in a specified MRI environment with specified conditions of use. Therefore, it is quite important to understand what kind of clinical phenomena would take place if an accidental scan of a conventional device was performed. If the MRI scanning of a patient with non MR compatible AIMD is needed from special medical requirement, intensive monitoring of heart activity during the scanning and repetitive threshold measurements after the scanning will be essential.
  • Keywords
    biomedical MRI; biomedical equipment; cardiology; patient monitoring; AIMD manufacturers; EMI phenomenon; MRI conditional devices; MRI scanning; accidental scan; active implantable medical devices; clinical phenomena; heart activity; intensive monitoring; repetitive threshold measurements; scanning threshold measurements; specified MRI environment; Coils; Electrodes; Heart; Magnetic fields; Magnetic resonance imaging; Pacemakers; Radio frequency; AIMD; MRI; magnet field gradient; rf pulses; static magnet field;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electromagnetic Compatibility, Tokyo (EMC'14/Tokyo), 2014 International Symposium on
  • Conference_Location
    Tokyo
  • Type

    conf

  • Filename
    6997180