• DocumentCode
    1791882
  • Title

    A CMOS liquid level to frequency converter with calibration circuits for detecting liquid level of intravenous drip

  • Author

    Cheng-Ta Chiang ; Ping-Chen Tsai

  • Author_Institution
    Dept. of Electr. Eng., Nat. Chia Yi Univ., Chiayi, Taiwan
  • fYear
    2014
  • fDate
    3-6 Aug. 2014
  • Firstpage
    342
  • Lastpage
    346
  • Abstract
    In this paper, a CMOS liquid level to frequency converter with calibration circuits for detecting liquid level of intravenous drip is newly proposed. A calibration technique is proposed in the work. Another innovation is that the outputs of the proposed chip are directly digitized, they could be easily sent over a wide range of transmission media, such as PSN, radio, optical, IR, ultrasonic, and etc. Before performing the proposed calibration circuits, the maximum linear error is 46.44% After calibration, the maximum linear error is reduced to 2.04% The liquid level range is 32 steps, and the corresponding output frequency range is 3.14 to 58.14 kHz. The chip area is 2 × 1.96 mm2. The proposed chip is suitable for devices of measuring liquid level, such as device of detecting liquid level of intravenous drip.
  • Keywords
    CMOS integrated circuits; biomedical equipment; calibration; frequency convertors; level measurement; level meters; measurement errors; CMOS liquid level to frequency converter; calibration circuit; frequency 3.14 kHz to 58.14 kHz; intravenous drip; linear error; liquid level detection; liquid level measurement; transmission media; CMOS integrated circuits; Calibration; Frequency conversion; Liquids; Magnetic liquids; Simulation; Transducers; Liquid level sensor; Liquid level transducer;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Mechatronics and Automation (ICMA), 2014 IEEE International Conference on
  • Conference_Location
    Tianjin
  • Print_ISBN
    978-1-4799-3978-7
  • Type

    conf

  • DOI
    10.1109/ICMA.2014.6885720
  • Filename
    6885720