Title :
Precision cavity control for the stable operation of a large ring laser gyroscope
Author :
Schreiber, K.U. ; Gebauer, A. ; Velikoseltsev, A. ; Wells, J.-P.R.
Author_Institution :
Forschungseinrichtung Satellitengeodaesie, Tech. Univ. Muenchen, Muenchen, Germany
fDate :
June 30 2014-July 4 2014
Abstract :
Currently the sensor performance of large ring laser gyroscopes is limited more by stability in the long term rather than measurement resolution. This is mostly because of a variable contribution of backscatter coupling between the two counter-propagating laser beams inside the square ring laser cavity. Introducing an atmospheric pressure stabilizing vessel around the ring laser structure allows us to compensate variations in the compression of the ring laser body by ambient pressure changes. Adding an interferometric feedback system takes this approach one step further in that it allows us to stabilize the length of the cavity to be stable to within 1 kHz of optical frequency. However it transpires that this precision cavity control is not sufficient to keep the backscatter coupling sufficiently constant to reduce the variation of the offset bias of the gyroscope to values of 10 μHz or below. A tightly controlled perimeter does not preclude small variations of the 4 individual sides of the gyroscope. In the absence of sufficient control over the backscatter process or sufficiently precise numerical estimate of the backscatter variation, a tight control of the length of the four individual arms of the gyroscope, in addition to the precision perimeter control appears to be a viable experimental approach. Monitoring the phase relationship between the ring laser beat note, taken at different corners of the gyroscope, provides the necessary access to the desired cavity control. This paper reports the first results of this investigation.
Keywords :
gyroscopes; laser beams; laser cavity resonators; laser feedback; laser frequency stability; measurement by laser beam; optical control; ring lasers; ambient pressure changes; atmospheric pressure stabilizing vessel; backscatter coupling; backscatter process; backscatter variation; cavity length; counter-propagating laser beams; frequency 1 kHz; frequency 10 muHz; interferometric feedback system; large ring laser gyroscopes; measurement resolution; offset bias variation; optical frequency; phase relationship; precision cavity control; precision perimeter control; pressure 1 atm; ring laser beat note; ring laser body compression; ring laser structure; sensor performance; square ring laser cavity; variable contribution; Backscatter; Cavity resonators; Gyroscopes; Laser stability; Laser theory; Measurement by laser beam; Ring lasers; Earth rotation; optical frequency stabilization; ring laser; space geodesy;
Conference_Titel :
Laser Optics, 2014 International Conference
Conference_Location :
Saint Petersburg
Print_ISBN :
978-1-4799-3884-1
DOI :
10.1109/LO.2014.6886320