• DocumentCode
    179886
  • Title

    Ultrasonic estimation of separation change between membranes in a human thorax phantom oriented to pleural effusion detection

  • Author

    Cruz-Prieto, J. ; Bazan, I. ; Ramirez-Garcia, A.

  • Author_Institution
    ESIME Zacatenco, Inst. Politec. Nac., Mexico City, Mexico
  • fYear
    2014
  • fDate
    Sept. 29 2014-Oct. 3 2014
  • Firstpage
    1
  • Lastpage
    6
  • Abstract
    The objective of this paper is to develop an ultrasound signal processing technique suitable to measure changes in distance between two membranes that simulate the pleural membranes in a basic human thorax phantom. The proposed technique is based on the correlation and autocorrelation methods. Obtaining an accuracy of 96.83 percent compared to those measures obtained with a caliper.
  • Keywords
    biological organs; biomedical ultrasonics; biomembranes; medical signal processing; phantoms; autocorrelation methods; basic human thorax phantom; caliper; pleural effusion detection; pleural membranes; separation change; ultrasonic estimation; ultrasound signal processing technique; Acoustics; Correlation; Electrical engineering; Estimation; Lungs; Phantoms; Ultrasonic variables measurement; A-Scan; autocorrelation; correlation; pleural effusion; ultrasound;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electrical Engineering, Computing Science and Automatic Control (CCE), 2014 11th International Conference on
  • Conference_Location
    Campeche
  • Print_ISBN
    978-1-4799-6228-0
  • Type

    conf

  • DOI
    10.1109/ICEEE.2014.6978275
  • Filename
    6978275