DocumentCode :
1802447
Title :
Pulse parameter dependence on transition position and epoch duration [pulse generators and sampling oscilloscopes]
Author :
Larson, Donald R. ; Paulter, Nicholas G., Jr. ; Bergman, David I.
Author_Institution :
Quantum Electr. Metrol. Div., NIST, Gaithersburg, MD, USA
Volume :
2
fYear :
2004
fDate :
18-20 May 2004
Firstpage :
1291
Abstract :
The pulse parameters of amplitude and transition duration are dependent on the epoch duration and transition occurrence instant of the pulse transition in the epoch. The primary explanations for the observed variations are the pulse aberrations and settling behavior of both the pulse generator and the measurement instrument (sampling oscilloscope). Measurement results are included for two pulse generators and two sampling oscilloscopes.
Keywords :
measurement errors; oscilloscopes; pulse generators; pulse measurement; pulse aberrations; pulse amplitude; pulse epoch duration; pulse generator settling behavior; pulse measurement instrument; pulse transition duration; pulse transition position parameter dependence; pulse variations; sampling oscilloscope; settling error; transition occurrence instant; Amplitude estimation; Attenuation; Bandwidth; Instruments; Metrology; NIST; Oscilloscopes; Pulse generation; Pulse measurements; Sampling methods;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Instrumentation and Measurement Technology Conference, 2004. IMTC 04. Proceedings of the 21st IEEE
ISSN :
1091-5281
Print_ISBN :
0-7803-8248-X
Type :
conf
DOI :
10.1109/IMTC.2004.1351302
Filename :
1351302