Properties and performance of rate detection algorithms in three implantable cardioverter-defibrillators
Author :
Olson, Walter H. ; Gunderson, Bruce D. ; Fang-Yen, Michael C. ; Hoyhtya, Adam K.
Author_Institution :
Medtronic Inc., Minneapolis, MN, USA
Abstract :
Three detection algorithms used in implantable cardioverter defibrillators (ICDs) are described and compared. Synthetic and prerecorded human tachyarrhythmias were applied to algorithm simulations. As rhythm variability increased, the consistency of the result near detection zone boundaries fell and had different trends for each algorithm. For 456 prerecorded rhythms and fixed zones, VT-VF sensitivities were 96-97% and specificities were 83-85%
Keywords :
defibrillators; electrocardiography; medical signal processing; algorithm simulations; detection zone boundaries; implantable cardioverter-defibrillators; prerecorded rhythms; rate detection algorithms performance; rate detection algorithms properties; rhythm variability; Cardiology; Computer aided manufacturing; Counting circuits; Detection algorithms; Event detection; Gain control; Humans; Medical treatment; Rhythm; USA Councils;