    Properties and performance of rate detection algorithms in three implantable cardioverter-defibrillators

  • Author

    Olson, Walter H. ; Gunderson, Bruce D. ; Fang-Yen, Michael C. ; Hoyhtya, Adam K.

  • Author_Institution
    Medtronic Inc., Minneapolis, MN, USA
  • fYear
    1994
  • fDate
    25-28 Sep 1994
  • Firstpage
    65
  • Lastpage
    68
  • Abstract
    Three detection algorithms used in implantable cardioverter defibrillators (ICDs) are described and compared. Synthetic and prerecorded human tachyarrhythmias were applied to algorithm simulations. As rhythm variability increased, the consistency of the result near detection zone boundaries fell and had different trends for each algorithm. For 456 prerecorded rhythms and fixed zones, VT-VF sensitivities were 96-97% and specificities were 83-85%
  • Keywords
    defibrillators; electrocardiography; medical signal processing; algorithm simulations; detection zone boundaries; implantable cardioverter-defibrillators; prerecorded rhythms; rate detection algorithms performance; rate detection algorithms properties; rhythm variability; Cardiology; Computer aided manufacturing; Counting circuits; Detection algorithms; Event detection; Gain control; Humans; Medical treatment; Rhythm; USA Councils;