Comparison of the performance of correlation waveform analysis in controlled environments
Author
Morris, MM ; Pariseau, BJ ; Jenkins, JM ; DiCarlo, LA
Author_Institution
Michigan Univ., Ann Arbor, MI, USA
fYear
1994
fDate
25-28 Sep 1994
Firstpage
57
Lastpage
60
Abstract
Maximizing the efficiency of antitachycardia devices (ATDs) requires minimizing their computational complexity. Reduction of the sampling rate of intracardiac signals favorably impacts ATD power and storage constraints as long as accurate arrhythmia detection is maintained. This study determined the upper and lower limits of acceptable bandpass filtering for an established morphometric method of intraventricular arrhythmia classification: correlation waveform analysis (CWA). CWA accurately discriminated sinus rhythm from ventricular tachycardia and ventricular fibrillation on an intra-patient but not an inter-patient basis. The optimum filter settings were 1-50 Hz for unipolar recorded data and 10-50 Hz for bipolar recorded data sampled at 120 Hz. In conclusion, morphometric analysis remains feasible with narrow bandpass data. Reduction in signal sampling rates with concomitant decreases in ATD computation and data storage is possible without adversely affecting the accuracy of intraventricular arrhythmia classification
Keywords
band-pass filters; electrocardiography; medical signal processing; waveform analysis; 1 to 120 Hz; acceptable bandpass filtering limits; antitachycardia devices efficiency maximization; bipolar recorded data; controlled environments; correlation waveform analysis performance; data storage; established morphometric method; intraventricular arrhythmia classification; narrow bandpass data; optimum filter settings; sinus rhythm; ventricular fibrillation; ventricular tachycardia; Algorithm design and analysis; Band pass filters; Electrodes; Fibrillation; Filtering; Rhythm; Sampling methods; Signal analysis; Signal processing algorithms; Signal sampling;
fLanguage
English
Publisher
ieee
Conference_Titel
Computers in Cardiology 1994
Conference_Location
Bethesda, MD
Print_ISBN
0-8186-6570-X
Type
conf
DOI
10.1109/CIC.1994.470250
Filename
470250
Link To Document