Title :
Pre- and postprocessing algorithms for the correction of position dependencies of image processing parameters in ultrasonographic prostate images
Author :
Huynen, A.L. ; Giesen, R.J.B. ; Aarnink, R.G. ; de la Rosette, J.J.M.C.H. ; Debruyne, F.M.J. ; Wijkstra, H.
Author_Institution :
Dept. of Urology, Univ. Hospital Nijmegen, Netherlands
Abstract :
Describes position dependent correction algorithms for image processing parameters in ultrasonographic images. The parameters are calculated using the co-occurrence algorithm. By the nature of ultrasound and the rotation of a single element transducer depth- and angle dependencies occur. For angle dependencies a pre-processing algorithm is described. For depth dependencies a post-calculation correction is described
Keywords :
biomedical ultrasonics; image classification; medical image processing; ultrasonic transducers; angle dependency; cancer detection; co-occurrence algorithm; depth dependency; image processing parameters; position dependencies; position dependent correction algorithms; post-calculation correction; postprocessing algorithms; pre-processing algorithm; preprocessing algorithms; single element transducer rotation; ultrasonographic prostate images; ultrasound; Biological materials; Biopsy; Image analysis; Image processing; Imaging phantoms; Physics; Probes; Ultrasonic imaging; Ultrasonic transducers; Visualization;
Conference_Titel :
Engineering in Medicine and Biology Society, 1994. Engineering Advances: New Opportunities for Biomedical Engineers. Proceedings of the 16th Annual International Conference of the IEEE
Conference_Location :
Baltimore, MD
Print_ISBN :
0-7803-2050-6
DOI :
10.1109/IEMBS.1994.411852