DocumentCode :
1816245
Title :
X-Ray CCD setup for large-area wire chamber calibration
Author :
Platzer, K. ; Duennweber, W. ; Wellenstein, H.
Author_Institution :
Phys. Dept., Univ. of Munich, Garching, Germany
Volume :
2
fYear :
2003
fDate :
19-25 Oct. 2003
Firstpage :
1083
Abstract :
A X-ray setup has been developed to measure wire positions of straw tube drift detectors for the COMPASS experiment. In an automatic scanning procedure wire positions can be determined with an accuracy better than 30 μm inside an area of 3.6 × 3.6 m2. This accuracy is achieved by an independent two step movement of CCD and X-ray source, combined with a large-distance calibration with the help of external reference wires. Within smaller distances of several centimeters the achieved accuracy is much higher, in the order of a few microns.
Keywords :
X-ray detection; charge-coupled devices; multiwire ionisation chambers; X-ray CCD setup; large-area wire chamber calibration; straw tube drift detectors; Calibration; Charge coupled devices; Detectors; Mesons; Particle beams; Physics; Protons; Telephony; Wire; X-ray imaging;