Title :
Usage of a capillary plate as a pre-gas-amplification device for neutron Micro-Strip Gas Chamber
Author :
Nakamura, T. ; Masaoka, S. ; Yamagishi, H. ; Sakasai, K. ; Soyama, K. ; Aizawa, K.
Author_Institution :
Japan Atomic Energy Res. Inst., Ibaraki, Japan
Abstract :
A capillary plate, an assembly of the fine glass-tubes, was applied as a pre-gas-amplification device for a neutron Micro-Strip Gas Chamber (MSGC) under high-gas pressure of helium-3. The authors have developed the method to evaluate the collection efficiency of the electrons between the capillary plate and MSGC at the neutron irradiation, and evaluated it as a function of the potential between them. The detector system comprised of the capillary plate and MSGC showed the collection efficiency of 17 % maximally at the tested configuration, and achieved the effective gas gain of ∼600 at the gas pressure of 3 atm with 10% mixture of ethane with helium-3. It was confirmed that the capillary plate could be used as a pre-gas-amplification device for neutron detection at a pressurized helium-3 gas.
Keywords :
multiwire ionisation chambers; neutron detection; capillary plate; fine glass-tubes; neutron Micro-Strip Gas Chamber; neutron irradiation; pre-gas-amplification device; Absorption; Assembly; Charge measurement; Current measurement; Detectors; Electrons; Gain measurement; Helium; Neutrons; Q measurement;
Conference_Titel :
Nuclear Science Symposium Conference Record, 2003 IEEE
Print_ISBN :
0-7803-8257-9
DOI :
10.1109/NSSMIC.2003.1351958