DocumentCode :
1820272
Title :
Micro-hole & strip plate detector for neutron detection
Author :
Veloso, J.F.C.A. ; Amaro, F. ; dos Santos, J.M.F. ; Mir, J.A. ; Derbyshire, G.E. ; Stenphenson, R. ; Rhodes, N.J. ; Schooneveld, E.M.
Author_Institution :
Dept. de Fisica, Aveiro Univ., Portugal
Volume :
1
fYear :
2003
fDate :
19-25 Oct. 2003
Firstpage :
257
Abstract :
We propose the use of argon-xenon penning-mixtures as the stopping gas, in substitution of carbon-based gases, in /sup 3/He neutron detectors. Such stopping gas presents a solution for the use of GEMS or MHSPs electron multipliers with suitable gains for neutron detection. With these mixtures it is possible to obtain a sealed detector with only rare-gas filling which is simple to purify and not subject to ageing. A MHSP gas detector filled with a 3-bar argon/50-mbar xenon/6-bar helium mixture can achieve gains above 2 × 10/sup 3/. This mixture allows neutron detection efficiencies of about 70% for a 2.5-cm thick absorption region and intrinsic position resolution (FWHM) of about 1.7-mm. Increasing the argon partial pressure to 6-bar, it is possible to reduce the detector intrinsic position resolution to 1-mm. The sensitivity to γ-rays will be the same when compared to that of 2.6-bar CF/sub 4/.
Keywords :
neutron detection; position sensitive particle detectors; Ar-Xe; Ar-Xe penning mixtures; electron multipliers; micro-hole detector; neutron detection; sealed detector; stopping gas; strip plate detector; Aging; Argon; Electron multipliers; Filling; Gas detectors; Gases; Helium; Neutrons; Strips; Xenon;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2003 IEEE
Conference_Location :
Portland, OR, USA
ISSN :
1082-3654
Print_ISBN :
0-7803-8257-9
Type :
conf
DOI :
10.1109/NSSMIC.2003.1352042
Filename :
1352042
Link To Document :
بازگشت