DocumentCode :
1820396
Title :
A parallel-plate resistive-anode gaseous detector for X-ray imaging
Author :
Khazins, D.M. ; Becker, B.L. ; Diawara, Y. ; Durst, R.D. ; He, B.B. ; Medved, S.A. ; Sedov, V. ; Thorson, T.A.
Author_Institution :
Bruker AXS Inc., Madison, WI, USA
Volume :
1
fYear :
2003
fDate :
19-25 Oct. 2003
Firstpage :
277
Abstract :
A novel gaseous detector, based on parallel-plate gas amplification with a resistive anode and remote readout electrode, is described. The detector is significantly less prone to discharges than other gaseous wireless detectors and unique in its ability to achieve high gain at high counting rates. A local counting rate > 5 × 105 counts/mm2-sec at a gas gain of up to 105 has been achieved in a 14 × 14 cm2 area detector for diffraction applications. The detector radiation hardness is characterized by the anode accumulated charge > 100 Coul/cm2. The detector is sealed and its re-gassing period is estimated to be 5-10 years. The operating characteristics of the new detector, its spatial resolution, and linearity are described and preliminary X-ray diffraction data are presented.
Keywords :
X-ray detection; X-ray imaging; avalanche photodiodes; photon counting; X-ray diffraction data; X-ray imaging; anode accumulated charge; detector radiation hardness; high counting rates; local counting rate; parallel-plate gas amplification; parallel-plate resistive-anode gaseous detector; re-gassing period; remote readout electrode; resistive anode; Anodes; Electrodes; Gas detectors; Insulation; Radiation detectors; Sparks; Strips; X-ray detection; X-ray detectors; X-ray imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2003 IEEE
ISSN :
1082-3654
Print_ISBN :
0-7803-8257-9
Type :
conf
DOI :
10.1109/NSSMIC.2003.1352046
Filename :
1352046
Link To Document :
بازگشت