• DocumentCode
    1821092
  • Title

    Noise and sampling analysis for multiple-image radiography

  • Author

    Marquet, Bruno ; Brankov, Jovan G. ; Wernick, Miles N.

  • Author_Institution
    Dept. of Electr. & Comput. Eng., Illinois Inst. of Technol., Chicago, IL
  • fYear
    2006
  • fDate
    6-9 April 2006
  • Firstpage
    1232
  • Lastpage
    1235
  • Abstract
    We have recently proposed a new X-ray imaging method called multiple-image radiography (MIR), which produces three images that separately depict the effects of X-ray absorption, X-ray refraction, and ultra-small-angle X-ray scattering. Because MIR is a relatively recent development, many of its basic properties have not yet been determined. In this paper we investigate the noise properties of MIR by deriving the Cramer-Rao lower bounds (CRLB) for the three MIR images. Next using this bound an evaluation of the uniform angular sampling acquisition procedure is given
  • Keywords
    X-ray absorption; X-ray scattering; diagnostic radiography; medical image processing; noise; Cramer-Rao lower bounds; X-ray absorption; X-ray imaging; X-ray refraction; multiple-image radiography; noise analysis; sampling analysis; ultra-small-angle X-ray scattering; uniform angular sampling acquisition; Biomedical engineering; Data acquisition; Diffraction; Electromagnetic wave absorption; Optical imaging; Protocols; Radiography; Sampling methods; X-ray imaging; X-ray scattering;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Imaging: Nano to Macro, 2006. 3rd IEEE International Symposium on
  • Conference_Location
    Arlington, VA
  • Print_ISBN
    0-7803-9576-X
  • Type

    conf

  • DOI
    10.1109/ISBI.2006.1625147
  • Filename
    1625147