DocumentCode
1821101
Title
Half-cone SPECT system for small animal imaging
Author
Li, Heng ; Zheng, Yibin ; Stolin, Alexander V. ; Kundu, Bijoy K. ; Williams, Mark B.
Author_Institution
Dept. of Electr. & Comput. Eng., Virginia Univ., Charlottesville, VA
fYear
2006
fDate
6-9 April 2006
Firstpage
1236
Lastpage
1239
Abstract
Molecular imaging studies using small animals such as mice require the development of high-sensitivity, high-resolution pinhole SPECT imaging systems. It is well known that the photon detection efficiency drops remarkably as the pinhole angle increases, and spatial resolution decreases as the magnification factor decreases. To obtain high efficiency and high resolution, the pinhole should be put as close as possible to the object while the pinhole angle remains as small as possible. To address these conflicting requirements, we proposed a half-cone imaging geometry in which the detector looks at only half of the FOV in each view, while the FOV being reconstructed remain the same. Our preliminary results demonstrated that SPECT system using this geometry could achieve higher sensitivity and better spatial resolution
Keywords
image reconstruction; image resolution; medical image processing; single photon emission computed tomography; half-cone SPECT system; high-sensitivity high-resolution pinhole SPECT imaging systems; image reconstruction; magnification factor; mice; molecular imaging; photon detection efficiency; pinhole angle; small animal imaging; spatial resolution; Animals; Apertures; Geometry; High-resolution imaging; Image reconstruction; Image resolution; Mice; Spatial resolution; X-ray detection; X-ray detectors;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Imaging: Nano to Macro, 2006. 3rd IEEE International Symposium on
Conference_Location
Arlington, VA
Print_ISBN
0-7803-9576-X
Type
conf
DOI
10.1109/ISBI.2006.1625148
Filename
1625148
Link To Document