• DocumentCode
    1823083
  • Title

    Development of a RadFET linear array for intracavitary in vivo dosimetry in external beam radiotherapy and brachytherapy

  • Author

    Price, Robert A. ; Benson, Chris ; Rodgers, Kenneth

  • Author_Institution
    Dept. of Phys., Clatterbridge Centre for Oncology NHS Trust, UK
  • Volume
    3
  • fYear
    2003
  • fDate
    19-25 Oct. 2003
  • Firstpage
    1494
  • Abstract
    We present the details of a new linear array dosimeter consisting of a chain of semiconductors mounted on an ultra-thin (50μm thick) flexible substrate and housed in an intracavitary catheter. The semiconductors, manufactured by NMKC Cork, have not been packaged and incorporate a passivation layer that allows them to be mounted on the substrate using flip-chip-bonding. This paper reports, for the first time, the construction of a multiple (ten) detector array suited to in vivo dosimetry in the rectum, esophagus and vagina during external beam radiotherapy, as well as being adaptable to in vivo dosimetry during brachytherapy and diagnostic radiology.
  • Keywords
    biological organs; brachytherapy; catheters; dosimeters; dosimetry; field effect transistors; flip-chip devices; radiology; semiconductor counters; 50 mum; RadFET linear array; brachytherapy; diagnostic radiology; esophagus; external beam radiotherapy; flip-chip-bonding; intracavitary catheter; intracavitary in vivo dosimetry; linear array dosimeter; multiple detector array; passivation layer; rectum; semiconductor detectors; ultra-thin flexible substrate; vagina; Adaptive arrays; Brachytherapy; Catheters; Dosimetry; In vivo; Passivation; Semiconductor device manufacture; Semiconductor device packaging; Structural beams; Substrates;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2003 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8257-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2003.1352160
  • Filename
    1352160