  • Abstract
    The development and initial evaluation of a high-resolution single photon emission tomography (SPECT) based system to image the biodistribution of radiolabeled tracers such as Tc-99m and I-125 in unrestrained/un-anesthetized mice. An infrared (IR) based position tracking apparatus has been developed and integrated into a SPECT gantry. The tracking system is designed to measure the spatial position of a mouse´s head at a rate of 10-15 frames per second with sub-millimeter accuracy. The high resolution, gamma imaging detectors are based on pixelated NaI(Tl) crystal scintillator arrays, arrays of compact position-sensitive photomultiplier tubes and novel readout circuitry for lower device cost while retaining high spatial resolution. Two SPECT gamma camera detector heads based on a 4 × 8 array of Hamamatsu R8520-C12 position sensitive photomultiplier tubes have been built and installed onto the gantry. The IR landmark-based pose measurement and tracking system is under development to provide animal position data during a SPECT scan. The animal position and orientation data acquired by the IR tracking system is used for motion correction during the tomographic image reconstruction.
  • Keywords
    biomechanics; gamma-ray detection; image motion analysis; image reconstruction; image resolution; medical image processing; optical tracking; single photon emission computed tomography; solid scintillation detectors; Hamamatsu R8520-C12 position sensitive photomultiplier tubes; IR landmark-based pose measurement; animal orientation; animal position; biodistribution; compact position-sensitive photomultiplier tubes; gamma imaging detectors; high-resolution single photon emission tomography; infrared based motion tracking; infrared based position tracking apparatus; motion correction; pixelated NaI(Tl) crystal scintillator arrays; radiolabeled tracers; readout circuitry; restraint free small animal SPECT imaging system; tomographic image reconstruction; unanesthetized mice; Animals; High-resolution imaging; Infrared imaging; Mice; Optical imaging; Position measurement; Sensor arrays; System testing; Tomography; Tracking;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2003 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8257-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2003.1352292
  • Filename
    1352292