Title :
FDG imaging of 1mm tumor with an ultra high resolution animal PET
Author :
Ishii, Kazuki ; Funaki, Y. ; Yamazaki, Hiroshi ; Matsuyama, S. ; Terakawa, Ayano ; Fujiwara, Masamichi ; Iwata, Ryosuke ; Kodama, Tomoya ; Watanabe, Yoshihiro ; Tanizaki, N. ; Amano, D. ; Yamaguchi, Toru
Author_Institution :
Dept. of Quantum Sci. & Energy Eng., Tohoku Univ., Sendai
Abstract :
Recently, we reported an animal semiconductor PET with the spatial resolution of 0.8 mm FWHM within the central 20 mm-diameter of FOV for the purpose of biomedical study using rats and mice. This ultra high spatial resolution was obtained by the use of small CdTe elements of 1.1mm x 1.0 mm x 5 mm. The FOV of this PET is 64 mm in diameter and 26 mm in axis. We applied to observe small tumors in mouse and succeeded to 1 Q obtain [ F]FDG images of mouse mammary carcinoma of ~1mm size.
Keywords :
biomedical imaging; cancer; image resolution; medical image processing; positron emission tomography; tumours; FDG imaging; mammary carcinoma; mice; rats; semiconductor; size 20 mm; spatial resolution; tumor; ultra high resolution animal PET; Animals; Biomedical engineering; High-resolution imaging; Image resolution; Mice; Neoplasms; Positron emission tomography; Power engineering and energy; Solid scintillation detectors; Spatial resolution; positron emission tomography;
Conference_Titel :
Biomedical Imaging: From Nano to Macro, 2008. ISBI 2008. 5th IEEE International Symposium on
Conference_Location :
Paris
Print_ISBN :
978-1-4244-2002-5
Electronic_ISBN :
978-1-4244-2003-2
DOI :
10.1109/ISBI.2008.4541315