    Combined patient motion compensation and dose rate variation to improve radiotherapy treatment dosimetric accuracy

  • Author

    Haas, Olivier C. L. ; Paluszczyszyn, Daniel ; Gillot, Guillaume ; Mills, J.A.

  • Author_Institution
    Control Theor. & Applic. Centre, Coventry Univ., Coventry, UK
  • fYear
    2010
  • fDate
    7-10 Sept. 2010
  • Firstpage
    1
  • Lastpage
    6
  • Abstract
    This paper demonstrates that using a novel combination of motion prediction algorithm, couch control system and couch nonlinearities compensator it is possible to realise a `breathing couch´ with a standard Elekta patient support system (PSS). The couch has been shown to be able to reduce the vertical and longitudinal motion of a surrogate located on a robotic arm to less than 1mm RMS and 2mm 95%CI. The irradiation of a moving target was compared to that of a fixed target and a moving target which motion is compensated by the PSS. No noticeable differences were observed between the films obtained for the fixed and the vertical & longitudinal compensated motions. When tracking is combined with dose rate adaption further improvement can be obtained without significantly increasing the treatment time.
  • Keywords
    dosimetry; manipulators; medical image processing; motion compensation; radiation therapy; video signal processing; 2D video tracking program; Elekta patient support system; PSS; couch control system; couch nonlinearities compensator; dose rate variation; fixed target; motion prediction algorithm; moving target; patient motion compensation; radiotherapy treatment dosimetric accuracy; robotic arm; measurements; medical robot; position accuracy; radiotherapy; tracking;
  • fLanguage
    English
  • Publisher
    iet
  • Conference_Titel
    Control 2010, UKACC International Conference on
  • Conference_Location
    Coventry
  • Electronic_ISBN
    978-1-84600-038-6
  • Type

    conf

  • DOI
    10.1049/ic.2010.0308
  • Filename
    6490766