Title :
Design for a high-resolution small-animal SPECT system using pixelated Si(Li) detectors for in vivo 125I imaging
Author :
Choong, W. -S ; Moses, W.W. ; Tindall, C.S. ; Luke, P.N.
Author_Institution :
Lawrence Berkeley Nat. Lab., CA, USA
Abstract :
We propose a design for a high-resolution SPECT system for in vivo 125I imaging in small animal using pixelated lithium-drifted silicon (Si(Li)) detectors. The detectors have high detection efficiency (>90%), good energy resolution (<15% FWHM), and good intrinsic spatial resolution (∼1 mm FWHM). The SPECT system has a dual head detector geometry with the distance between the detectors ranging 30-50 mm to allow for the shortest imaging distance between the mouse and the detectors. The detectors, each with an active area of 64 mm×40 mm (64×40 array of 1 mm2 pixels and a 6 mm thick Si(Li) detector), are mounted on a rotating gantry with an axial field-of-view of 64 mm. Using a high-resolution parallel-hole collimator, the expected spatial resolution is 1.6 mm FWHM at an imaging distance of 20 mm, and sensitivity is 6.7 cps/μCi. 125I is a readily available radioisotope with a long half-life of 59.4 days and it is commonly used to label biological compounds in molecular biology. Conventional gamma cameras are not optimized to detect the low emission energies (27 to 35 keV) of 125I. However, Si(Li) detector provides an ideal solution for detecting the low-energy emissions of 125I. In addition, a SPECT system based on Si(Li) detector can potentially be less expensive than other detector technology because of its ease of operation and the availability of many resources for processing silicon. This will increase the availability of small-animal SPECT systems in biomedical research, thereby providing an important tool to facilitate the research in mouse biology.
Keywords :
biomedical imaging; collimators; gamma-ray detection; iodine; molecular biophysics; position sensitive particle detectors; radioactive tracers; radioisotopes; sensitivity; silicon radiation detectors; single photon emission computed tomography; 27 keV to 35 KeV; 30 to 50 mm; 6 mm; 64 mm; 125I; active area; axial field-of-view; biological compounds; biomedical research; dual head detector geometry; energy resolution; gamma cameras; half-life; high detection efficiency; high-resolution parallel-hole collimator; high-resolution small-animal SPECT system; imaging distance; in vivo 125I imaging; intrinsic spatial resolution; low-energy emission; molecular biology; mouse biology; pixelated Si(Li) detectors; pixelated lithium-drifted silicon detectors; radioisotope; rotating gantry; small animal; Animals; Availability; Gamma ray detection; Gamma ray detectors; High-resolution imaging; In vivo; Mice; Pixel; Silicon; Spatial resolution;
Conference_Titel :
Nuclear Science Symposium Conference Record, 2003 IEEE
Print_ISBN :
0-7803-8257-9
DOI :
10.1109/NSSMIC.2003.1352360