Title :
X-ray dose savings with reduced acquisition rate pulsed fluoroscopy for cardiac interventional procedures
Author :
Xue, Ping ; Aufrichtig, Richard ; Wilson, David L.
Author_Institution :
Dept. of Biomed. Eng., Case Western Reserve Univ., Cleveland, OH, USA
Abstract :
Based upon threshold detectability of low-contrast cylinders, the authors estimate X-ray dose savings for pulsed fluoroscopy at 15 acq/sec as compared to 30 acq/sec. The authors use an adaptive four-alternative forced-choice experiment, they find dose savings from 7-50%. Savings depend upon cylinder size and velocity
Keywords :
cardiology; diagnostic radiography; dosimetry; 4-alternative forced-choice experiment; X-ray dose savings; cardiac interventional procedures; cylinder size; low-contrast cylinders; medical diagnostic imaging; reduced acquisition rate pulsed fluoroscopy; threshold detectability; Background noise; Biomedical engineering; Catheters; Displays; Engine cylinders; Imaging phantoms; Performance evaluation; X-ray detection; X-ray detectors; X-ray imaging;
Conference_Titel :
Engineering in Medicine and Biology Society, 1994. Engineering Advances: New Opportunities for Biomedical Engineers. Proceedings of the 16th Annual International Conference of the IEEE
Conference_Location :
Baltimore, MD
Print_ISBN :
0-7803-2050-6
DOI :
10.1109/IEMBS.1994.411922