    X-ray dose savings with reduced acquisition rate pulsed fluoroscopy for cardiac interventional procedures

  • Author

    Xue, Ping ; Aufrichtig, Richard ; Wilson, David L.

  • Author_Institution
    Dept. of Biomed. Eng., Case Western Reserve Univ., Cleveland, OH, USA
  • fYear
    1994
  • fDate
    3-6 Nov 1994
  • Firstpage
    524
  • Abstract
    Based upon threshold detectability of low-contrast cylinders, the authors estimate X-ray dose savings for pulsed fluoroscopy at 15 acq/sec as compared to 30 acq/sec. The authors use an adaptive four-alternative forced-choice experiment, they find dose savings from 7-50%. Savings depend upon cylinder size and velocity
  • Keywords
    cardiology; diagnostic radiography; dosimetry; 4-alternative forced-choice experiment; X-ray dose savings; cardiac interventional procedures; cylinder size; low-contrast cylinders; medical diagnostic imaging; reduced acquisition rate pulsed fluoroscopy; threshold detectability; Background noise; Biomedical engineering; Catheters; Displays; Engine cylinders; Imaging phantoms; Performance evaluation; X-ray detection; X-ray detectors; X-ray imaging;