• DocumentCode
    1828623
  • Title

    Homogenization of the MultiChannel PM gain by inserting light attenuating masks

  • Author

    Christ, D. ; Hollendung, A. ; Larue, H. ; Parl, C. ; Streun, M. ; Weber, S. ; Ziemons, K. ; Halling, H.

  • Author_Institution
    Zentrallabor fuer Elektronik, Forschungszentrum Juelich GmbH, Germany
  • Volume
    4
  • fYear
    2003
  • fDate
    19-25 Oct. 2003
  • Firstpage
    2382
  • Abstract
    MultiChannel Photomultipliers (PM), like the R7600-00-M64 or R5900-00-M64 from Hamamatsu, are often chosen as photodetectors in high-resolution positron emission tomography (PET). A major problem of this PM is the nonuniform channel gain. In order to solve this problem, light attenuating masks were created. The aim of the masks is a homogenization of the output of all 64 channels using different hole sizes at the channel positions. The hole area, which is individually defined for the different channels, is inversely proportional to the channel gain. The measurements by inserting light attenuating masks improved a homogenization to a ratio of 1:1.2.
  • Keywords
    optical attenuators; photodetectors; photomultipliers; positron emission tomography; readout electronics; scintillation counters; Hamamatsu; MultiChannel Photomultipliers; PET; R5900-00-M64; R7600-00-M64; channel positions; high-resolution positron emission tomography; hole area; hole sizes; homogenization; light attenuating masks; multichannel PM gain; nonuniform channel gain; photodetectors; readout; Anodes; Attenuation measurement; Frequency measurement; Gain measurement; Light emitting diodes; Performance gain; Photomultipliers; Positron emission tomography; Pulse measurements; Spatial resolution;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2003 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8257-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2003.1352375
  • Filename
    1352375