DocumentCode :
1829183
Title :
Performance of 256ch flat panel PS-PMT with small crystals for a DOI PET detector
Author :
Inadama, Naoko ; Murayama, Hideo ; Watanabe, Mitsuo ; Omura, Tomohide ; Yamashita, Takaji ; Kawai, Hideyuki ; Orita, Narimichi ; Tsuda, Tomoaki
Author_Institution :
Nat. Inst. of Radiol. Sci., Chiba, Japan
Volume :
4
fYear :
2003
fDate :
19-25 Oct. 2003
Firstpage :
2486
Abstract :
A 256ch flat panel position sensitive photomultiplier tube (256ch FP-PMT) was evaluated for application to a depth of interaction (DOI) PET detector composed of small crystal elements. The 256ch FP-PMT has 52 mm×52 mm opening area with 89% useful area. Its 256 anodes are placed at a 3.04 mm interval. To estimate the FP-PMT performance, light spread functions of anodes at central and peripheral regions of a prototype 256ch FP-PMT were measured with a Gd2SiO5 (GSO) crystal sized 1.42 mm×1.42 mm×4.5 mm, one quarter of the area of the FP-PMT anode area. It was found that light spread functions of peripheral anodes were similar to the functions of central anodes and also the functions measured with a 2.9 mm×2.9 mm×7.5 mm GSO crystal, almost the same size as the anode area. Transit time fluctuation among the 256ch FP-PMT anodes was estimated by measuring time resolutions of a LuYSiO5 (LYSO) crystal at some central and peripheral positions. The LYSO crystal has the same dimensions as the smaller GSO crystal BaF2 was used for a reference detector. The obtained distributions show about the same time resolution between the central and peripheral positions and the full width at half maximum is (366±15) ps. The transit time fluctuation is ±106 ps. Two-dimensional position histograms of a 32×32 GSO crystal array and two layers of 9×9 GSO crystal arrays are obtained by uniform gamma-ray irradiation. The crystal dimensions are 1.42 mm×1.42 mm×4.5 mm. The 32×32 array covers all the useful area. The 9×9×2 array is coupled to peripheral region. The resultant histograms confirm that a 256ch FP-PMT has enough capability for identifying crystals of this size even on the periphery.
Keywords :
gamma-ray effects; photomultipliers; position sensitive particle detectors; positron emission tomography; solid scintillation detectors; BaF2; DOI PET detector; FP-PMT anode area; Gd2SiO5 crystal; LuYSiO5 crystal; central anodes; crystal elements; depth of interaction PET detector; flat panel PS-PMT; flat panel position sensitive photomultiplier tube; full width at half maximum; gamma-ray irradiation; light spread functions; peripheral anodes; transit time fluctuation; two-dimensional position histograms; Anodes; Area measurement; Crystals; Fluctuations; Histograms; Photomultipliers; Position sensitive particle detectors; Positron emission tomography; Prototypes; Size measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2003 IEEE
ISSN :
1082-3654
Print_ISBN :
0-7803-8257-9
Type :
conf
DOI :
10.1109/NSSMIC.2003.1352397
Filename :
1352397
Link To Document :
بازگشت