• DocumentCode
    1829672
  • Title

    Effect of crystal shape, size and reflector type on operation characteristics of gamma-radiation detectors based on CsI(Tl) and CsI(Na) scintillators

  • Author

    Globus, M.E. ; Grinyov, B.V. ; Ratner, M.A.

  • Author_Institution
    Inst. for Single Crystals, Kharkov, Ukraine
  • Volume
    2
  • fYear
    1996
  • fDate
    2-9 Nov 1996
  • Firstpage
    724
  • Abstract
    Operation characteristics of CsI(Tl) and CsI(Na) scintillation detectors, to a large degree connected with light collection in crystals, are calculated for various shapes, sizes and reflecting surface types. Allowance is made for the true light reflection indicatrix which is characterized by the effective mirror constituent of the reflected light, p. Its value ⟨p⟩, averaged over incidence angle, is used for the classification of reflecting surfaces. Operation characteristics (in particular, spectrometric ones) are found to be essentially dependent on ⟨p⟩ Tables of operation characteristics, given below, permit one to make inferential conclusions on an optimal combination of the shape, sizes and the reflecting surface version
  • Keywords
    caesium compounds; light reflection; sodium; solid scintillation detectors; thallium; CsI(Na); CsI(Tl); CsI:Na; CsI:Tl; crystal shape; effective mirror constituent; gamma-radiation detectors; inferential conclusions; light collection; light reflection indicatrix; operation characteristics; reflected light; reflecting surface version; reflector type; Crystals; Energy resolution; Gamma ray detection; Gamma ray detectors; Mirrors; Optical reflection; Rough surfaces; Shape; Spectroscopy; Surface roughness;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE
  • Conference_Location
    Anaheim, CA
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-3534-1
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1996.591420
  • Filename
    591420