Title :
Motion artefact reduction for exact cone-beam CT reconstruction algorithms
Author :
Koken, P. ; Köhler, Th ; Bontus, C. ; Proksa, R.
Author_Institution :
Philips Res. Lab., Hamburg, Germany
Volume :
4
fYear :
2003
fDate :
19-25 Oct. 2003
Firstpage :
2985
Abstract :
Katsevich proposed a reconstruction algorithm for helical cone-beam CT which is theoretically exact. Even for large cone angles, it performs very well on static objects. But due to its limitation to the PI illumination window, it produces strong artefacts in the presence of motion. We propose a method to reduce these spreading motion artefacts significantly by adding a correction image. It uses a small portion of the voxel-dependent over-scan. Results of a simulation study with a 128 row scanner are shown.
Keywords :
computerised tomography; image motion analysis; image reconstruction; medical image processing; Katsevich reconstruction; PI illumination window; cone angles; correction image; helical cone-beam CT reconstruction algorithms; motion artefact reduction; spreading motion artefacts; voxel-dependent over-scan; Computed tomography; Detectors; Filtering; Focusing; Image quality; Image reconstruction; Laboratories; Lighting; Reconstruction algorithms; Thorax;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2003 IEEE