• DocumentCode
    1834995
  • Title

    Improved Waveguide Calibration of Vector Network Analyzers in WR15

  • Author

    Gianfortune, P. ; Radmanesh, M.

  • Author_Institution
    Maury Microwave Corporation
  • Volume
    15
  • fYear
    1989
  • fDate
    32660
  • Firstpage
    112
  • Lastpage
    132
  • Abstract
    This paper discusses factors affecting the accuracy of Vector Network Analysis in millimeter waveguides. Among these are accuracy improvements provided by a new precision WR15 waveguide flange which maintains compatibility with the current UG-385/U flanges. The new wR15 waveguide flange design offers much improved performance in the frequency range of 50 to 75 GHz. Measurement results for flange repeatability along with accuracy results for residual directivity and source match, both for Maury and UG-385/U standard flanges, are presented.
  • Keywords
    Calibration; Frequency;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    ARFTG Conference Digest-Spring, 33rd
  • Conference_Location
    Long Beach, CA, USA
  • Print_ISBN
    0-7803-5686-1
  • Type

    conf

  • DOI
    10.1109/ARFTG.1989.323943
  • Filename
    4119509