DocumentCode
1838059
Title
Planar Resistors for Probe Station Calibration
Author
Walker, D.K. ; Williams, D.F. ; Morgan, J.M.
Author_Institution
National Institute of Standards and Technology, Boulder, Colorado
Volume
22
fYear
1992
fDate
Dec. 1992
Firstpage
1
Lastpage
9
Abstract
This paper investigates the effects of variations in sheet resistance, geometry, distance from the probe tip, and fabrication processes on the impedance of planar nickel-chromium resistors. Resistor reactance is a strong function of film resistance, but depends only weakly on geometry and distance from the probe tip. Photoresist contamination in the resistive film induces more complicated impedance behavior, even at low frequencies. The impact on circuit design and time- and frequency-domain calibrations is considered in light of these results.
Keywords
Calibration; Circuit synthesis; Contamination; Fabrication; Frequency; Geometry; Impedance; Probes; Resistors; Resists;
fLanguage
English
Publisher
ieee
Conference_Titel
ARFTG Conference Digest-Fall, 40th
Conference_Location
Orlando, FL, USA
Print_ISBN
0-7803-5686-1
Type
conf
DOI
10.1109/ARFTG.1992.326993
Filename
4119651
Link To Document