DocumentCode
1838306
Title
High resolution PET detector using wavelength shifting optical fibers and microchannel plate PMT with delay line readout
Author
Williams, M.B. ; Sealock, R.M. ; Majewski, S. ; Weisenberger, A.G.
Volume
2
fYear
1996
fDate
2-9 Nov 1996
Firstpage
1228
Abstract
We are developing a PET detector with depth of interaction (DOI) sensitivity for breast and small animal imaging. The detector uses layers of high density scintillators sandwiched between orthogonal ribbons of wavelength shifting optical fibers. Fibers are coupled to microchannel plate (MCP) based, position sensitive photomultiplier tubes (PSPMTs) containing a crossed delay line readout anode. Preliminary measurements have been made using blue-green waveshifting fiber ribbons coupled to LSO. The measured photoelectron yield due to wavelength shifted light from the fibers is sufficient to ensure a Poisson probability of only 1.0% that no photoelectron is generated in a given MCP PSPMT. Measurement of internally reflected scintillation light permits energy resolution of better than 25% at 511 keV. Optical fiber readout of high density inorganic crystals provides significantly larger detective area per unit phototube area than do one-to-one coupling schemes using PSPMTs. The use of the MCP PSPMT solves problems that have prevented other types of PSPMTs from being used in such a configuration, namely poor resolution at the low light output level of the optical fibers
Keywords
diagnostic radiography; fibre optic sensors; microchannel plates; optical delay lines; photomultipliers; position sensitive particle detectors; positron emission tomography; solid scintillation detectors; 511 keV; Lu2(SiO4)O; Poisson probability; blue-green waveshifting fiber ribbons; breast imaging; crossed delay line readout anode; delay line readout; depth of interaction sensitivity; high density scintillators; high resolution PET detector; microchannel plate PMT; orthogonal ribbons; photoelectron yield; position sensitive photomultiplier tubes; positron emission tomography; small animal imaging; wavelength shifting optical fibers; Animals; Breast; Detectors; High-resolution imaging; Microchannel; Optical fibers; Optical imaging; Optical sensors; Positron emission tomography; Wavelength measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE
Conference_Location
Anaheim, CA
ISSN
1082-3654
Print_ISBN
0-7803-3534-1
Type
conf
DOI
10.1109/NSSMIC.1996.591631
Filename
591631
Link To Document