Save The "Thru" in the A.N.A. Calibration
Author
Ferrero, A. ; Pisani, U. ; Sanpietro, F.
Author_Institution
POLITECNICO DI TORINO, Corso Duca degli Abruzzi 24, 10129 TORINO, ITALY, Tel. xx39-11-5644047 Fax: xx39-11-564-4099
Volume
22
fYear
1992
fDate
3-4 Dec. 1992
Firstpage
128
Lastpage
135
Abstract
The conventional network analyzer (NWA) two-port calibration procedures require a standard thru line to be connected between the ports. Unfortunately in many applications, for example when measuring MMIC or on-wafer devices with not aligned ports, a custom thru line must be used. The procedure here applied overcomes the difficulty due to the poor knowledge of this thru element since it is based on a generic reciprocal unknown two port structure, provided that its S21 phase shift is roughly known. Some experimental comparisons with other well sound calibration techniques will be here presented where different reciprocal two-port structures were used as unknown thru.
Keywords
Calibration; Coaxial components; Equations; Intelligent networks; MMICs; Measurement standards; Scattering parameters; System testing; Transmission line matrix methods;