DocumentCode
1839314
Title
Verification of Commercial Probe-Tip Calibrations
Author
Marks, Roger B. ; Williams, Dylan F.
Author_Institution
National Institute of Standards and Technology, Mail Code 813.06,325 Broadway, Boulder, CO 80303
Volume
24
fYear
1993
fDate
Dec. 1993
Firstpage
37
Lastpage
44
Abstract
We present results of a verification procedure useful in evaluating the accuracy of probe-tip scattering parameter measurements. The procedure was applied to calibrations and measurements performed in industrial laboratories. Actual measurement discrepancies, due primarily to calibration errors, are directly com pared to bounds determined by the comparison method. The results demonstrate the utility of the verification technique as well as serious flaws, particularly at high frequencies, in some conventional calibrations.
Keywords
Calibration; Coplanar waveguides; Electromagnetic measurements; Gallium arsenide; Impedance; Measurement uncertainty; NIST; Performance evaluation; Scattering parameters; Testing;
fLanguage
English
Publisher
ieee
Conference_Titel
ARFTG Conference Digest-Fall, 42nd
Conference_Location
San Jose, CA, USA
Print_ISBN
0-7803-5686-1
Type
conf
DOI
10.1109/ARFTG.1993.327039
Filename
4119710
Link To Document