Abstract :
A vertical coil array is designed to increase the signal-to-noise ratio (SNR) for magnetic resonance imaging (MRI). The SNR increase is realized by properly combining the output of each coil element of the array, and in this way the coupling within the array is totally removed. The method of determining the combiner coefficients is introduced in a simulated MRI system and the SNR is shown to increase with the number of coils in the array.
Keywords :
biomedical MRI; coils; image resolution; MRI; SNR; coil element; coupling; high quality image resolution; magnetic resonance imaging; signal-to-noise ratio; vertical coil array; Arrays; Coils; Magnetic resonance imaging; Mutual coupling; Signal to noise ratio; Vectors; Vertical coil array; combiner coefficients; magnetic resonance imaging; signal-to-noise ratio (SNR);