• DocumentCode
    1842133
  • Title

    FNS control system tuning and clinical ratings of performance

  • Author

    Abbas, James J. ; Chizeck, Howard J.

  • Author_Institution
    Cleveland Veterans Adm. Med. Center, OH, USA
  • fYear
    1989
  • fDate
    9-12 Nov 1989
  • Firstpage
    1494
  • Abstract
    A feedback controller of coronal plane hip angle was implemented in paraplegic subjects using functional neuromuscular stimulation (FNS). In an evaluation based on standard temporal response properties, the feedback-controlled system exhibited substantial improvements when compared to an open-loop system. An evaluation based on subjective clinical observations of the experimental trials indicates that standard control system design criteria may not correspond to desired performance characteristics. Consideration of the clinical ratings and identified system model parameters suggests other design criteria which may be more relevant
  • Keywords
    biocontrol; muscle; neurophysiology; orthotics; clinical ratings; coronal plane hip angle; design criteria; feedback controller; feedback-controlled system; functional neuromuscular stimulation; open-loop system; paraplegic subjects; standard temporal response properties; system model parameters; Adaptive control; Biomedical measurements; Control systems; Delay estimation; Error correction; Feedback control; Hip; Mathematical model; Parameter estimation; Steady-state;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1989. Images of the Twenty-First Century., Proceedings of the Annual International Conference of the IEEE Engineering in
  • Conference_Location
    Seattle, WA
  • Type

    conf

  • DOI
    10.1109/IEMBS.1989.96306
  • Filename
    96306