    Calibration Comparison Method for Vector Network Analyzers

  • Author

    Marks, Roger B. ; Jargon, Jeffrey A. ; Juroshek, John R.

  • Author_Institution
    National Institute of Standards and Technology, 325 Broadway, Mail Code 813.06, Boulder, CO 80303 USA, Tel: (303)497-3037 Fax: (303)497-3970, E-mail: marks@nist.gov
  • Volume
    30
  • fYear
    1996
  • fDate
    Dec. 1996
  • Firstpage
    38
  • Lastpage
    45
  • Abstract
    We present a technique for comparing the scattering parameter measurements made with respect to two vector network analyzer calibrations. This method determines the worst-case measurement error bounds on any calibration from a benchmark calibration, assuming the two are similar to first order. We illustrate our method by examining the differences between an open-short-load-thru (OSLT) and a sliding load calibration, both of which are available commercially on a variety of vector network analyzers.
  • Keywords
    Calibration; Code standards; Electronic mail; Error correction; Instruments; Measurement errors; Measurement standards; Postal services; Scattering parameters; Testing;