• DocumentCode
    1845284
  • Title

    An Automatic Calibration Approach For Left Ventricular Volume Assessment

  • Author

    Beatrice, P. ; Regis, V. ; Philippe, D. ; Remy, P. ; Francoise, P. ; Gerard, F.

  • Author_Institution
    Univ. Lyon 1, Lyon
  • fYear
    2007
  • fDate
    22-26 Aug. 2007
  • Firstpage
    4460
  • Lastpage
    4463
  • Abstract
    The development of interventional procedures in the catheterization laboratory drives the introduction of simple, fast and accurate quantification methods. Quantification is used to evaluate the volume of the left ventricle but also for choosing the size of a balloon or a stent during the percutaneous coronary angioplasty procedure. The objective of the calibration procedure is to establish the relation between the true size of the analyzed organ and its apparent size in the image. This calibration step aims at measuring the pixel size in mm. Traditional approach measures a known object on the image, but suffers from several limitations and constraints; in particular, the size of the reference object is often not exactly known. Besides, the therapeutic procedures are done in the continuity of the diagnosis and the physician has a very little amount of time to spend on the determination of the quantitative information like the chosen reference size. A new approach has been proposed, allowed by the introduction of the flat-panel technology and modern positioning systems. With the absence of distortion in the image and accurate knowledge of the system geometry, a model can be used to automatically identify the calibration factor using an additional input provided by the operator: the target organ height above the tabletop. The objective of our study was to estimate this input parameter in the particular case of the Left Ventricle (LV) volume angiographic computation. In this article, we compared two estimates of the LV center height and assessed their performances through the computation of the LV volume. We concluded on the reliability of the fully automatic calibration procedure.
  • Keywords
    angiocardiography; blood vessels; calibration; cardiovascular system; surgery; angiographic computation; automatic calibration approach; catheterization; flat-panel technology; interventional procedures; left ventricular volume assessment; percutaneous coronary angioplasty; pixel size measurement; positioning systems; therapeutic procedures; Angioplasty; Calibration; Catheterization; Geometry; Image analysis; Laboratories; Parameter estimation; Particle measurements; Size measurement; Solid modeling; Angiography; Calibration; Female; Heart Ventricles; Humans; Male; Models, Biological; Organ Size;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2007. EMBS 2007. 29th Annual International Conference of the IEEE
  • Conference_Location
    Lyon
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-0787-3
  • Type

    conf

  • DOI
    10.1109/IEMBS.2007.4353329
  • Filename
    4353329