    A simple and precise charge balancing method for voltage mode stimulation

  • Author

    Ranjandish, R. ; Shoaei, O.

  • Author_Institution
    Sch. of Electr. & Comput. Eng., Univ. of Tehran, Tehran, Iran
  • fYear
    2014
  • fDate
    22-24 Oct. 2014
  • Firstpage
    376
  • Lastpage
    379
  • Abstract
    This paper proposes a simple charge balancing method for voltage mode stimulation which can precisely balance the injected charge in order to avoid electrode dissolution or tissue damage using high-frequency voltage pulses. By explaining the tissue model, the effect of such a high-frequency pulses is investigated. Additionally, a prototype of the system is implemented by employing low power off-the-shelf discrete components. The experimental results show the proper performance of the proposed system.
  • Keywords
    bioelectric phenomena; biological tissues; high-voltage techniques; low-power electronics; physiological models; prosthetics; electrode dissolution; high-frequency voltage pulses; injected charge; low power off-the-shelf discrete components; precise charge balancing method; simple charge balancing method; tissue damage; tissue model; voltage mode stimulation; Charge transfer; Computational modeling; Educational institutions; Electric potential; Electrodes; Integrated circuit modeling; Safety; Electrical stimulation; charge balancing; implantable medical devices; voltage mode stimulation;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Circuits and Systems Conference (BioCAS), 2014 IEEE
  • Conference_Location
    Lausanne
  • Type

    conf

  • DOI
    10.1109/BioCAS.2014.6981741
  • Filename
    6981741