Abstract :
A new type of two-dimensional position sensitive gas counter for transmission counter with a backgammon-shaped flat electrode has been developed. The two dimensional position read-out can be achieved with the combination of a backgammon-shaped flat electrode read-out as horizontal position and the drift time measurement read-out as vertical position. The gas counter is designed as a transmission trajectory counter that can be used to determine angle of incident charged particles in charged particle detecting system. To check the characteristics of this gas counter, test experiments are performed with coincidence plastic scintillation counter and β-ray source of 90Sr
Keywords :
beta-ray detection; drift chambers; position sensitive particle detectors; β-ray source; backgammon-drift read-out technique; backgammon-shaped flat electrode; coincidence plastic scintillation counter; incident charged particles; two-dimensional position sensitive gas counter; two-dimensional transmission counter; Counting circuits; Electrodes; Performance evaluation; Plastics; Radiation detectors; Scintillation counters; Solid scintillation detectors; Strontium; Testing; Time measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference Record, 1995., 1995 IEEE
Conference_Location :
San Francisco, CA
Print_ISBN :
0-7803-3180-X
Type :
conf
DOI :
10.1109/NSSMIC.1995.504198
Filename :
504198
Link To Document :
بازگشت