A targeted sparse readout for multi-anode photo-multipliers and optically isolated crystals
Author
Clancy, R.L. ; Thompson, C.J. ; Robar, J.L. ; Murthy, K. ; Beuville, E. ; Moses, W.W.
Author_Institution
Neurological Inst., McGill Univ., Montreal, Que., Canada
Volume
1
fYear
1995
fDate
21-28 Oct 1995
Firstpage
409
Abstract
Multi anode photo-multipliers are able to localize the position of a light source, and measure its intensity. Normally these devices have their anode wires connected to a resistor network. If the light originates in an optically collimated, pixilated scintillation crystal array, only a few of the anode wires receive any significant signal. If only those wires whose signal is above the noise threshold are used to localize the light, the signal-to-noise ratio improves (same signal but less noise). We found that only about 6 wires on each axis have a significant signal when a 511 keV gamma ray is detected in a 2 mm×2 mm BGO crystal. Using a custom made integrated circuit which provides the binary code of the address corresponding to the biggest of 16 analog inputs, and a sample-and-hold on each anode wire we hope to direct a six channel ADC to sample the anode wires which provide the best positioning signal on each axis of the PMT. The shape of the signal envelope permits identification of crystals which would otherwise be beyond the PMT´s field of view. This readout should enhance these PMTs in application like positron emission mammography, and small animal PET imaging
Keywords
application specific integrated circuits; biomedical equipment; detector circuits; diagnostic radiography; gamma-ray detection; nuclear electronics; photomultipliers; position sensitive particle detectors; positron emission tomography; solid scintillation detectors; 2 mm; 51.1 keV; BGO crystal; PET imaging; analog inputs; binary code; custom made integrated circuit; intensity; light source; multianode photomultipliers; noise threshold; optically collimated pixilated scintillation crystal array; optically isolated crystals; positioning signal; positron emission mammography; sample-and-hold; signal-to-noise ratio; six channel ADC; targeted sparse readout; Anodes; Integrated circuit noise; Light sources; Optical arrays; Optical collimators; Optical noise; Position measurement; Resistors; Signal to noise ratio; Wires;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium and Medical Imaging Conference Record, 1995., 1995 IEEE
Conference_Location
San Francisco, CA
Print_ISBN
0-7803-3180-X
Type
conf
DOI
10.1109/NSSMIC.1995.504255
Filename
504255
Link To Document