Title
Automatic amplitude adjustment of the pacemaker output voltage
Author
Fröhlich, R. ; Bolz, A. ; Hardt, R. ; Hubmann, M. ; Schaldach, M.
Author_Institution
Dept. for Biomed. Eng., Friedrich-Alexander-Univ., Erlangen, Germany
fYear
1994
fDate
3-6 Nov 1994
Firstpage
55
Abstract
Subthreshold pacing may threaten the life of the patient. In order to avoid this risk, the pacemaker should be able to check the effectiveness of its pacing pulse. Therefore, the aim of this work was to develop an algorithm for automatic amplitude adjustment (AAA), which is based on the measurement of the ventricular evoked response (VER). In clinical investigations the AAA for the VVI-mode was tested using an external system, consisting of an external programmable pacemaker (EDP 30 A, BIOTRONIK) and a laptop. The algorithm has been checked during 5 pacemaker implantations. The results show, that the system detects “no capture” reliably. During each investigation the threshold was measured automatically. The measured threshold voltage was reproducible equal to the value determined by the physician before
Keywords
biocontrol; pacemakers; voltage control; BIOTRONIK; VVI-mode; automatic amplitude adjustment algorithm; external programmable pacemaker; measured threshold voltage; no capture detection; pacemaker implantations; pacemaker output voltage; pacing pulse effectiveness checking; patient life threat; ventricular evoked response measurement; Biomedical engineering; Biomedical measurements; Frequency; Myocardium; Pacemakers; Portable computers; Pulse measurements; System testing; Threshold voltage; Voltage measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1994. Engineering Advances: New Opportunities for Biomedical Engineers. Proceedings of the 16th Annual International Conference of the IEEE
Conference_Location
Baltimore, MD
Print_ISBN
0-7803-2050-6
Type
conf
DOI
10.1109/IEMBS.1994.412107
Filename
412107
Link To Document