Title : 
Frequency biases in a cold-atom coherent population trapping clock
         
        
            Author : 
Donley, E.A. ; Blanshan, E. ; Esnault, Francois-Xavier ; Kitching, J.
         
        
            Author_Institution : 
NIST, Boulder, CO, USA
         
        
        
        
        
        
            Abstract : 
A compact cold-atom clock based on coherent population trapping (CPT) has been developed. The clock typically demonstrates a short-term fractional frequency stability of 4×10-11/√τ, limited by frequency noise on the interrogation lasers. The clock interrogates a sample of 2×105 atoms under free-fall with typical cycle and Ramsey periods of 50 ms and 16 ms, respectively. The largest two systematic frequency shifts that can limit the clock´s long-term stability are the light shift and the Doppler shift.
         
        
            Keywords : 
Doppler shift; atomic clocks; laser frequency stability; optical saturable absorption; CPT; clock interrogation; coherent population trapping; fractional frequency stability; frequency bias; frequency noise; interrogation laser; Atomic beams; Atomic clocks; Couplings; Doppler shift; Laser stability; NIST;
         
        
        
        
            Conference_Titel : 
Frequency Control Symposium (FCS), 2014 IEEE International
         
        
            Conference_Location : 
Taipei
         
        
        
            DOI : 
10.1109/FCS.2014.6859921