• DocumentCode
    186135
  • Title

    Arduino-based portable system for bioelectrical impedance measurement

  • Author

    Corbellini, Simone ; Vallan, Alberto

  • Author_Institution
    Dipt. di Elettron. e Telecomun., Politec. di Torino, Turin, Italy
  • fYear
    2014
  • fDate
    11-12 June 2014
  • Firstpage
    1
  • Lastpage
    5
  • Abstract
    Electrochemical Impedance Spectroscopy (EIS) has been recently proposed as a simple non-invasive technique to monitor the amount of fat and liquids contained inside the human body. While the technique capabilities are still questioned, a simple and low cost device capable of performing this kind of measurements would help testing it on many patients with minimal effort. This paper describes an extremely low cost implementation of an EIS system suitable for medical applications that is based on a simple commercial Arduino Board whose cost is below 50$. The circuit takes advantage of the ADC and DAC made available by the microcontroller of the Arduino boards and employs a logarithmic amplifier to extend the impedance measuring range to 6 decades without using complex programmable gain amplifiers. This way the device can use electrodes with sizes in the range of 1 cm2 to 40 cm2. The EIS can be measured in the frequency range of 1 Hz to 100 kHz and in the impedance range of 1 kΩ to 1 GΩ. The instrument automatically compensates the DC voltage due to the skin/electrode contact and runs on batteries. The EIS traces can be stored inside the device and transferred to the PC with a wireless link avoiding safety issues.
  • Keywords
    bioelectric phenomena; biomedical measurement; electric impedance measurement; electrochemical impedance spectroscopy; portable instruments; ADC; Arduino-based portable system; DAC; EIS; bioelectrical impedance measurement; electrochemical impedance spectroscopy; frequency 1 Hz to 100 kHz; microcontroller; Current measurement; Electrodes; Gain measurement; Impedance; Impedance measurement; Microcontrollers; Performance evaluation; Body Fat; Edema; Electroimpedance Impedance measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Medical Measurements and Applications (MeMeA), 2014 IEEE International Symposium on
  • Conference_Location
    Lisboa
  • Print_ISBN
    978-1-4799-2920-7
  • Type

    conf

  • DOI
    10.1109/MeMeA.2014.6860044
  • Filename
    6860044