Abstract :
Integration of a 2D ultrasound transducer array with front-end signal conditioning electronics poses significant fabrication and packaging challenges due to the required number of elements and small element pitch. In this paper, we demonstrate a CMUT-in-CMOS approach that produces ultrasonic transducers and circuits on a single chip using a standard foundry CMOS process, augmented only with two blanket post-process steps for sacrificial etching and cavity sealing. 2D arrays as large as 16 times 16, with integrated on-chip buffers, switches and multiplexers were fabricated and successfully characterized. This approach provides a straightforward path to commercially viable 2D arrays for 3D medical imaging.
Keywords :
CMOS integrated circuits; ultrasonic transducer arrays; 2D ultrasound transducer array; 3D medical imaging; CMUT-in-CMOS ultrasonic transducer arrays; cavity sealing; front-end signal conditioning electronics; integrated on-chip buffers; on-chip electronics; sacrificial etching; standard foundry CMOS process; CMOS process; Electronics packaging; Etching; Fabrication; Foundries; Switches; Switching circuits; Ultrasonic imaging; Ultrasonic transducer arrays; Ultrasonic transducers; 3D imaging; CMUT; Ultrasound; transducer array;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Solid-State Sensors, Actuators and Microsystems Conference, 2009. TRANSDUCERS 2009. International
Conference_Location :
Denver, CO
Print_ISBN :
978-1-4244-4190-7
Electronic_ISBN :
978-1-4244-4193-8
Type :
conf
DOI :
10.1109/SENSOR.2009.5285878
Filename :
5285878
Link To Document :
بازگشت