Pulsations in the blood pressure cuff: Oscillations or arterial pulses?
Author
Jilek, Jiri ; Stork, Milan
Author_Institution
Carditech, Culver City, CA, USA
fYear
2011
fDate
7-8 Sept. 2011
Firstpage
1
Lastpage
4
Abstract
Most automatic blood pressure monitors use the so-called oscillometric method. The method evaluates amplitudes of arterial pulses elicited in the cuff during blood pressure measurement. These pulses are commonly called “oscillations”. The objective of this study was to show that the contours of pulses in the cuff are arterial waveforms rather than oscillations. An experimental computer-based system for data acquisition and processing was used to acquire and process wrist cuff pulses and finger photoplethysmographic waveforms. The wrist cuff pulse contours were compared to radial artery pulse contours acquired by other methods. The comparison revealed that wrist cuff waveforms acquired at or below diastolic pressure exhibit properties of radial artery waveforms as obtained by other invasive or noninvasive methods. New terms “cuff-pulse method” and “cuff-arterial waveforms” were proposed.
Keywords
blood pressure measurement; blood vessels; data acquisition; medical computing; oscillations; patient monitoring; photoplethysmography; arterial pulses; arterial waveforms; automatic blood pressure monitors; blood pressure cuff; blood pressure measurement; computer-based system; cuff-arterial waveform; cuff-pulse method; data acquisition; diastolic pressure; finger photoplethysmographic waveforms; oscillations; oscillometric method; radial artery waveforms; wrist cuff pulse contours; wrist cuff pulses; Arteries; Biomedical monitoring; Blood pressure; Fingers; Oscillators; Pregnancy; Wrist;
fLanguage
English
Publisher
ieee
Conference_Titel
Applied Electronics (AE), 2011 International Conference on
Conference_Location
Pilsen
ISSN
1803-7232
Print_ISBN
978-1-4577-0315-7
Electronic_ISBN
1803-7232
Type
conf
Filename
6049042
Link To Document