• DocumentCode
    1879263
  • Title

    Relations Between Accuracy of Vessel Measurement and Physical Image Quality Characteristics in Multi-detector CT Scanner

  • Author

    Azuma, Y. ; Akagawa, T. ; Hamaguchi, H. ; Goto, S. ; Sumimoto, T.

  • Author_Institution
    Fac. of Health Sci., Okayama Univ. Med. Sch.
  • fYear
    2006
  • fDate
    24-27 April 2006
  • Firstpage
    438
  • Lastpage
    442
  • Abstract
    The influence of image quality characteristics on the accuracy of vessel measurement was investigated using simple vascular models, the comparison of the profile curve of the cross section of the vascular models and physical image quality characteristics such as modulation transfer function (MTF), Wiener spectrum (WS) and contrast, with changing scan conditions. The profile curves in the section part that greatly influences the measurement accuracy of the vessel diameter has changed by X-ray tube voltage. The noise has increased evenly in all frequency band regions at the same time though the contrast rose in 2 and 4 mm vascular models. The change was not seen in MTF. It can be said that the decision of the scan parameters that raises the contrast and decreases the noise will improve the measurement accuracy of the vessel diameter
  • Keywords
    X-ray tubes; biomedical measurement; computerised tomography; optical scanners; Wiener spectrum; X-ray tube voltage; measurement accuracy; modulation transfer function; multidetector CT scanner; physical image quality characteristics; profile curve; vascular model cross section; vessel measurement; Attenuation; Computed tomography; Containers; Detectors; Image quality; Imaging phantoms; Plastics; Sensor arrays; Steel; Transfer functions; Diameter of a vascular model; Image quality characteristics; Measurement accuracy of vessels; Multi-detector CT scanner; Profile curve;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 2006. IMTC 2006. Proceedings of the IEEE
  • Conference_Location
    Sorrento
  • ISSN
    1091-5281
  • Print_ISBN
    0-7803-9359-7
  • Electronic_ISBN
    1091-5281
  • Type

    conf

  • DOI
    10.1109/IMTC.2006.328512
  • Filename
    4124361