• DocumentCode
    18919
  • Title

    Improvements in the Low-Frequency Characteristic and Sensitivity of a Thin-Film Multijunction Thermal Converter in Vacuum

  • Author

    Amagai, Yasutaka ; Fujiki, Hiroyuki ; Shimizume, Koji ; Kishino, Kaname ; Hidaka, Shigeru

  • Author_Institution
    Nat. Metrol. Inst. of Japan, Nat. Inst. of Adv. Ind. Sci. & Technol., Tsukuba, Japan
  • Volume
    64
  • Issue
    6
  • fYear
    2015
  • fDate
    Jun-15
  • Firstpage
    1570
  • Lastpage
    1575
  • Abstract
    We have improved the ac-dc transfer differences and enhanced the sensitivity in our thin-film multijunction thermal converters (TFMJTCs) in vacuum. The ac-dc transfer differences of the TFMJTCs are less than 1 μV/V down to 1 Hz and 10 μV/V down to 0.2 Hz owing to the increased thermal time constant. The measured output voltage exceeds 120 mV at 10-4 Pa owing to the reduction of heat conduction to the air. The sensitivity of the device, defined as the ratio of the output voltage to the input power, is 2.4 V/W, which is approximately three times larger than that measured at ambient pressure. Numerical simulations of the heat transfer of our TFMJTC explain the experimental results well.
  • Keywords
    AC-DC power convertors; heat conduction; thermocouples; thin films; ac-dc transfer differences; frequency 0.2 Hz; frequency 1 Hz; heat conduction; heat transfer; low-frequency characteristic; thermal time constant; thin-film multijunction thermal converter; Frequency measurement; Heat transfer; Heating; Pressure measurement; Sensitivity; Temperature measurement; Voltage measurement; AC-DC difference; electrical variable measurement; numerical analysis; thermal converter; thermal voltage converter; thermal voltage converter.;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2014.2385133
  • Filename
    7010042