DocumentCode
1893833
Title
Piezoelectric micro cantilever sensor for non-labeling detection of biomarker
Author
Lee, Yeolho ; Seo, Hyejung ; Jeon, Sangmin ; Moon, Wonkyu
Author_Institution
Dept. of Mech. Eng., Pohang Univ. of Sci. & Technol. (POSTECH), Pohang
fYear
2008
fDate
26-29 Oct. 2008
Firstpage
250
Lastpage
253
Abstract
Using the piezoelectric micro cantilever sensor, biomarkers of various diseases are detected at their critical concentration or below. The detection principle of the sensor is the mass-micro-balancing technique that utilizes the change of the mechanical resonance frequency due to an object attached on a structure. An Ohm-contact electrode is installed to remove the parasitic capacitance effects between the electrodes in the sensor. The first resonant frequency of the piezoelectric micro cantilever can be monitored by measurement of the conductance component of the admittance. The results of the bio-marker-detection experiments show that the detection sensitivity of the micro cantilever is from 2 to 3 orders of magnitude higher than that of the conventional quartz-crystal micro-balancing detection.
Keywords
bioMEMS; biosensors; cantilevers; crystal resonators; electric admittance measurement; microbalances; piezoelectric oscillations; Ohm contact electrode; conductance measurement; mass microbalancing technique; mechanical resonance frequency change; microcantilever first resonant frequency; nonlabeled biomarker detection; parasitic capacitance effects removal; piezoelectric microcantilever sensor; Biomarkers; Biosensors; Capacitive sensors; Electrodes; Mechanical sensors; Object detection; Parasitic capacitance; Parasitic diseases; Resonance; Resonant frequency;
fLanguage
English
Publisher
ieee
Conference_Titel
Sensors, 2008 IEEE
Conference_Location
Lecce
ISSN
1930-0395
Print_ISBN
978-1-4244-2580-8
Electronic_ISBN
1930-0395
Type
conf
DOI
10.1109/ICSENS.2008.4716428
Filename
4716428
Link To Document