Title :
2-port calibration without a through connection using 1-port switched loads
Author :
Haynes, Mark ; Moghaddam, Mahta
Author_Institution :
Dept. of Electr. Eng. & Comput. Sci., Univ. of Michigan, Ann Arbor, MI, USA
Abstract :
Many instances occur in microwave measurements where a 2-port calibration is required but a physical through connection is not available. An example is in antenna range measurements, where S21 is sought but it is impossible to perform a complete 2-port calibration. Another example is calibrating out the effects of cable motion in mechanically automated measurements systems, where manually performing a 2-port calibration for each position is impractical. In this paper, we derive a procedure to determine the 2-port S-matrix of an error circuit from 1-port measurements using three non-standard loads. We then derive calibration procedures to remove this error network in both 1 and 2 port scenarios and demonstrate their effectiveness in bench tests.
Keywords :
S-matrix theory; calibration; microwave measurement; 1-port switched loads; 2-port S-matrix; 2-port calibration; automated measurements systems; microwave measurements; Antenna measurements; Attenuators; Calibration; Measurement uncertainty; Motion measurement; Reflection; Switches;
Conference_Titel :
Antennas and Propagation Society International Symposium (APSURSI), 2010 IEEE
Conference_Location :
Toronto, ON
Print_ISBN :
978-1-4244-4967-5
DOI :
10.1109/APS.2010.5562155