• DocumentCode
    1913686
  • Title

    A Novel Method of Measurement of Oxygen Saturation in Arterial Blood

  • Author

    Reddy, K. Ashoka ; George, Boby ; Mohan, N. Madhu ; Kumar, V. Jagadeesh

  • Author_Institution
    Dept. of Electr. Eng., IIT Madras, Chennai
  • fYear
    2008
  • fDate
    12-15 May 2008
  • Firstpage
    1627
  • Lastpage
    1630
  • Abstract
    Existing methods of pulse oximetry rely on a calibration curve for the estimation of oxygen saturation in arterial blood (SpO2). A novel method of computation of SpO2 that does not require the use of a calibration curve is presented in this paper. Based on a model proposed here for the propagation of light through skin, tissue, bone and blood, suitable processing steps are identified so that the analytical expression derived for the estimation of SpO2 becomes free of patient and sensor dependent parameters. Experimental results presented herein establish the efficacy of the proposed method.
  • Keywords
    biomedical measurement; blood vessels; light propagation; oxygen; arterial blood; light propagation; oxygen saturation measurement; pulse oximetry; Blood; Calibration; Fingers; Infrared detectors; Instrumentation and measurement; Light sources; Optical propagation; Oxygen; Pulse measurements; Wavelength measurement; Bio-medical signal processing; Oxygen saturation; Photoplethysmograph; Pulse oximeter;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference Proceedings, 2008. IMTC 2008. IEEE
  • Conference_Location
    Victoria, BC
  • ISSN
    1091-5281
  • Print_ISBN
    978-1-4244-1540-3
  • Electronic_ISBN
    1091-5281
  • Type

    conf

  • DOI
    10.1109/IMTC.2008.4547304
  • Filename
    4547304