• DocumentCode
    1925044
  • Title

    PET transmission scanning with a variable speed orbiting rod source

  • Author

    Leaf-Lensmire, Karen J. ; Stearns, Charles W. ; Colsher, James G.

  • Author_Institution
    GE Medical Syst., Milwaukee, WI, USA
  • fYear
    1992
  • fDate
    25-31 Oct 1992
  • Firstpage
    979
  • Abstract
    The authors have constructed a model which makes it possible to simulate the action of a variable-speed ORS (orbiting rod source) and to form variance sinograms for attenuation-corrected emission scans. The variance sinograms are then reconstructed and image variance estimates are obtained. Simulated results from the GE 2048 PET (positron emission tomography) scanner and experimental results form the GE Advance scanner were used to verify the model. It is shown that for noncircular objects the variance at the image center can be reduced by adopting a variable speed source orbit. In simulation, images derived from attenuation-corrected emission data also demonstrated improved uniformity. In experiment, the statistical uniformity of the variable speed ORS was improved in some areas of centralized anatomy while not in others
  • Keywords
    computerised tomography; radioisotope scanning and imaging; GE 2048 PET scanner; GE Advance scanner; PET transmission scanning; centralized anatomy; image center; medical diagnostic imaging; nuclear medicine; positron emission tomography; statistical uniformity; variable speed orbiting rod source; variance sinograms; Attenuation; Detectors; Filters; Geometry; Image reconstruction; Imaging phantoms; Positron emission tomography; Scattering; Solid modeling; Tellurium;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference, 1992., Conference Record of the 1992 IEEE
  • Conference_Location
    Orlando, FL
  • Print_ISBN
    0-7803-0884-0
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1992.301086
  • Filename
    301086